Vici D-2-I Manual de usuario

Busca en linea o descarga Manual de usuario para Sensores Vici D-2-I. VICI D-2-I User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 37
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Valco Instruments Co. Inc.
Pulsed Discharge
Detector
Models D-2 and D-2-I
Instruction Manual
Rev 2/14
Valco Instruments Co. Inc.
800 · 367· 8424 sales
713
· 688· 9345 tech
713
· 688· 8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
· 41 · 925· 6200 phone
Int + 41
· 41 · 925· 6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact::
Vista de pagina 0
1 2 3 4 5 6 ... 36 37

Indice de contenidos

Pagina 1 - Instruction Manual

Valco Instruments Co. Inc.Pulsed Discharge Detector Models D-2 and D-2-I Instruction Manual Rev 2/14 Valco Instruments Co. Inc.800 · 367· 8424 sales

Pagina 2

6IntroductionPDECDPDHIDMODE120160200804024028032036040010245122561286432168421ATTENUATIONDETECTOR TEMP.DISCHARGEZEROV ECDnA HIDPULSED DISCHARGE CONTRO

Pagina 3 - Table of Contents

7TEMPERATURE control and indicator Sets the temperature (°C) of the detector heater block. The indicator is steadily on when maximum power is being a

Pagina 4

8Display Indicates standing current in the PDHID mode or bias voltage in the PDECD mode, auto-scaled to the most suitable range. In the un-zeroed PDH

Pagina 5 - Introduction

9IntroductionSpecicationsMains (line): 115/230 V~50/60 Hz, 175 VA Fuse: 2 A, time-delay, 5 x 20 mm Pressure: 6.9 kPa (1 psi) operating, 6.9

Pagina 6 - Electron Capture (PDECD)

10System RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high purity grade ga

Pagina 7 - Safety Notes and Information

11System RequirementsGas SpecicationsFor Model D-2Detector ModePDECD PDHID Ar-PDPID Kr-PDPID Xe-PDPID Discharge gasHelium Helium 2% Ar in He 1.5% Kr

Pagina 8 - Description of Controls

12GC Column SelectionTo prevent detector contamination, we strongly recom-mend disconnecting the column from the detector dur-ing column bakeout proce

Pagina 9

13InstallationThe detector is usually mounted on top of the GC column oven. The cabling as supplied requires the discharge module to be located withi

Pagina 10 - PULSED DISCHARGE CONTROLLER

14Horizontal MountingSome older GCs have access to the column oven through the side of the GC. This does not present a problem as far as operation of

Pagina 11

15Installing and Purging the Gas Regulator1. Make sure the on/off valve on the helium cylinder is completely closed. Screw the CGA fitting nut of the

Pagina 12

This page intentionally left blank for printing purposes

Pagina 13 - Specications

16Installing and Purging the Helium Purier1. If the pressure regulator has a 1/8" male cone-type outlet port, install the Valco 1/8" exter

Pagina 14 - System Requirements

17Column ConnectionThe following applies only to systems with the PD-D2 detector. On the PD-D2-IS detector, the column is connected to the column in

Pagina 15 - Gas Specications

18ECD Mode0.00.53.51.01.52.02.53.0Peak height (nA)Length of column in detector (cm)9.2 9.4 9.6 10.84.09.8 10.0 10.2 10.4 10.6Detector temperature: 50

Pagina 16 - GC Column Selection

19InstallationElectrical ConnectionsCAUTION: Do not use a wrench to tighten the SMC connectors on the bias and electrometer cables. Connections sho

Pagina 17 - Installation

20InstallationInitial Power-UpNOTE: This section refers to the D-2 operating in the HID mode and to the D-2-I. For instructions on using the D-2 in

Pagina 18 - DOPANT VENT

21InstallationTroubleshooting High Background CurrentIf the background current does not drop below 2 nA even after a 12 hour bakeout, there is either

Pagina 19

22Column Bakeout PrecautionsTo prevent detector contamination, we strongly recommend disconnecting the column from the detector during column bakeout

Pagina 20

23Mode Selection and SetupHelium Ionization ModeSince the PDHID mode provides a better indication of the cleanliness and the integrity (leak tightness

Pagina 21 - Testing for Leaks

24Mode Selection and Setupamounts of pure helium and pure dopant through a tee. Since all gas streams must pass through a Valco purifier, the second o

Pagina 22 - HID Mode

25Electron Capture Mode (PD-D2 only)Since the PDHID mode provides a better indication of the cleanliness and the integrity (leak-tightness) of the sys

Pagina 23 - Electrical Connections

Table of ContentsIntroduction Description and Operating Principles ...1 Safety Notes an

Pagina 24 - Initial Power-Up

2610. If the bias voltage is not 1 V or less with a set current of 10 nA, the dopant flow rate may need to be adjusted. The optimum flow rate is de

Pagina 25

27ECD ModeDetector temperature: 50 CDischarge gas flow-rate: 32 mL/min0.1Background bias voltage V0 (V)Dopant Flow-rate (3% Xe, mL/min)0.0 7.01011.0 2

Pagina 26 - Column Bakeout Precautions

2801.8Bias voltage V0 (V)Detector temperature (C)050 100 150 200 250 300 3500.20.40.60.81.01.21.41.6061Peak height (V)050 3502345Detector temperature

Pagina 27 - Mode Selection and Setup

29Appendix: PD-C2 Controller (shipping February 2014)If you have been directed to this part of the manual, you are probably replacing an older PD-C2

Pagina 28

30UNATTENUATED OUTPUTS Normally connected to a data acquisition system or other recording means. The two unattenuated outputs are 0 - 10V full scale

Pagina 29

31  

Pagina 30

32WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyer may also have other rights that vary from state to state. For a per

Pagina 31

33Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below, (see Initial Power-Up, page 20), the standin

Pagina 32 - Peak height (V)

This page intentionally left blank for printing purposes

Pagina 33 - (shipping February 2014)

1IntroductionNOTE: The contents of this manual apply to either the HID/ECD dual mode detection system (D-2) or the HID single mode detection system (

Pagina 34

2Helium Ionization (PDHID)The PDHID is essentially non-destructive (0.01 - 0.1% ionization) and highly sensitive. The response to organic compounds i

Pagina 35 - Note About Compatibility

3dynamic range whereas, in the D-2, a unique system of feedback to the bias electrode is employed. It operates by comparing the detector output curre

Pagina 36 - Warranty

4mains (line) plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by t

Pagina 37 - Detector Performance Log

5Description Quantity Product numberDetector cell, PDHID 1 PD-D2-ISPulse supply module 1 PD-M2 Includes:Cable, contr

Modelos relacionados D-2

Comentarios a estos manuales

Sin comentarios