Vici D-4-I-TQ-R Manual de usuario

Busca en linea o descarga Manual de usuario para Sensores Vici D-4-I-TQ-R. VICI D-4-I-TQ-R User Manual [en] Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 22
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Valco Instruments Co. Inc.
Pulsed Discharge Detector
Models D-4-I-TQ-R
and D-4-I-TQI-R
Instruction Manual
Rev 1/15
Valco Instruments Co. Inc.
800 · 367· 8424 sales
713
· 688· 9345 tech
713
· 688· 8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
· 41 · 925· 6200 phone
Int + 41
· 41 · 925· 6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact::
Vista de pagina 0
1 2 3 4 5 6 ... 21 22

Indice de contenidos

Pagina 1 - Instruction Manual

Valco Instruments Co. Inc.Pulsed Discharge Detector Models D-4-I-TQ-R and D-4-I-TQI-R Instruction Manual Rev 1/15 Valco Instruments Co. Inc.800 · 3

Pagina 2

6InstallationGeneral Precautions • Do not use plastic/polymer or copper tubes for gas handling andinterconnectons. Use only stainless steel tubing w

Pagina 3 - Table of Contents

7InstallationGAS CHROMATOGRAPHTEE(ZT1)HELIUMPURIFIERDISCHARGE GAS(99.999% He)TGA-R-30F60PRESTRICTOR(30 mL/min minimum)VENTCOLUMNINLETDISCHARGE GASINLE

Pagina 4

8Installation7. Open the cylinder valve to pressurize the regulator once again. Close thevalve and observe the needle of the high pressure gauge for

Pagina 5 - Introduction

9Detector Electrical ConnectionsCAUTION: Do not use a wrench to tighten the SMCconnectors on the bias and electrometer cables.Connections should be fi

Pagina 6 - Safety Notes and Information

10Initial Power-UpCAUTION: Always make sure that discharge gas isflowing before heating and powering up the detector.1. Before installing the column,

Pagina 7

11InstallationCapillary Column ConnectionIf the capillary column adapter is installed in the column inlet:1. Make a mark on the column 14.5 cm from th

Pagina 8 - The discharge gas

123. Set aside the 1/8" wrench and completely remove the adapter from thecolumn inlet.4. Screw the packed column adapter into the column inlet by

Pagina 9 - Gas Specifications

13Mode Selection and SetupHelium Ionization ModeIf the instructions at the top of page 11 were properly executed, the columnshould already be properly

Pagina 10 - Installation

14Mode Selection and SetupAr-PDPIDChanging the discharge gas from helium to a mixture of 2% argon inhelium changes the photon energy level from the 17

Pagina 11 - Gas Connections

15TroubleshootingHigh Background CurrentLow SensitivityPurple/bluePinkDisconnect the columnand cap the column inletBaseline current is still high(>

Pagina 12

This page intentionally left blank for printing purposes

Pagina 13

16High Noise LevelIf the noise lievel is high:1. See if it improves with the GC fan turned off.2. Check the column insertion position (14.5 cm for cap

Pagina 14 - Initial Power-Up

17WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Pagina 15 - Packed Column Connection

18Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up,page 10), the standing

Pagina 16 - Exercise

Table of ContentsIntroduction Description and Operating Principles ...1 Safety Notes an

Pagina 17 - Mode Selection and Setup

This page intentionally left blank for printing purposes

Pagina 18

1IntroductionDescription and Operating PrincipleThe PD-D-4-I-TQI-R and PD-D-4-I-TQ-R are non-radioactive pulsed dischargeionization detectors (PDIDs)

Pagina 19 - Troubleshooting

2Helium Ionization (PDHID) ModeThe PDHID is essentially non-destructive (0.01 - 0.1% ionization) and highlysensitive. The response to organic compoun

Pagina 20 - High Noise Level

3Components of the Detector SystemComponents of the detector systems are listed in the tables below. Check the package contents to verify that everyt

Pagina 21 - Warranty

4IntroductionSystem RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high puri

Pagina 22 - Detector Performance Log

5IntroductionedoMrotceteDDIHDPDIPDP-rADIPDP-rKDIPDP-eXsagegrahcsiDmuileHeHnirA%2eHnirK%5.1eHnieX%8.0sagreirraCmuileH*****Gas Specifications** Any gas

Modelos relacionados D-4-I-TQI-R

Comentarios a estos manuales