Vici D-4-I-TQ-R Manual de usuario Pagina 22

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 22
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 21
18
Detector Performance Log
In addition to the occasions indicated in the Comments area of the table below (see
Initial Power-Up
,
page 10), the standing current should be observed and logged after any system change. Logging
the standing current (with and without the column) on a regular basis is also an effective monitor of
system integrity (leaktightness and cleanliness). To check sensitivity continuity, we recommend
tracking the internal standard (quantity on column/area count). Additional log pages can be down-
loaded from the support/manuals section of www.vici.com.
Date
Detector Model: Serial Number: Date of purchase:
Operator Column
temp
(°C)
Detector
temp
(°C)
Standing
current
(nA)
Noise
level
SampleComments
Initial power-up (IPU) 100——
IPU, detector to analysis temp
IPU, column installed Ambient
IPU, column to analysis temp
D-4-I-TQ-R
Vista de pagina 21
1 2 ... 17 18 19 20 21 22

Comentarios a estos manuales

Sin comentarios