Setting the Chamber Temperature via Serial Port (RS-232)
Refer to the section titled “Serial Port Communications” on page 20.
Installing the Permeation Device(s)
1. Turn off the heater power switch.
2. With the tool provided, rotate the panel lock screw 90° counterclockwise.
3. Gently pull the front panel assembly out.
CAUTION: The permeation chamber cap may be warm to
touch. If the calibrator has been in use, avoid exposure to
gas vapors while opening the chamber cap by using appro-
priate mitigation and personal protection equipment.
4. Unscrew the chamber cap with the tool provided.
5. Add or remove the permeation device(s) with the supplied forceps or other
tool appropriate to the job.
6. Secure the chamber cap with the tool provided..
7. Run a pressure/decay test to make sure that the chamber cap is leak tight.
8. Reinstall the front panel assembly and rotate the panel lock screw
90° clockwise.
9. Turn on the heater power switch and allow one hour for equilibration. Refer
to the separate instructions accompanying the permeation tube to make sure
that the selected temperature is compatible with the permeation tube being
used.
The Dynacalibrator is ready to be put into service.
Important Permeation Device Considerations
• Ifmorethanonetubeistobeusedintheovenatthesametime,orderallthe
tubes with permeation rates given at the same temperature.
• Certiedpermeationdevicesshouldbeusedonlyatthetemperaturespecied
on the certificate.
• Usinganypermeationdevicebeyonditsrecommendedtemperaturerange
could result in the destruction of the device by explosion and/or changes
in the membrane characteristics. If in doubt, contact VICI Metronics or their
authorized representative with the part number of the device to determine its
maximum temperature limit.