Vici 340 Dynacalibrator Manual de usuario Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
31
Configuration 5
in this configuration (Figure 10), pump P1 draws gas in through the supply
inlet on the back panel and delivers it to accumulator AC1. The accumulator
smoothes out flow fluctuations generated by the pumping action and provides
a port for releasing pump backpressure through pressure relief valve RV1, set at
5 psig. The tee at the output of the accumulator splits the supply into carrier and
dilution streams, each of which is plumbed to a back-panel outlet port which
must be plugged if an external supply is used. These ports can be connected
individually through external scrubbing media (filters, driers, etc.) or simply
looped to their respective inlet ports. Filters F1 and F2 remove particulates
larger than 5 microns from these two streams.
©










§








©







©

Advanced Theory of Operation
Figure 10: Diagram of Configuration 5
Vista de pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 43 44

Comentarios a estos manuales

Sin comentarios