Vici TCD2 Manual de usuario

Busca en linea o descarga Manual de usuario para Sensores Vici TCD2. VICI TCD2 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 25
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Valco Instruments Co. Inc.
tcd2.p65
Rev 4/12
Printed in USA
Microvolume Thermal
Conductivity Detector
Instruction Manual
Valco Instruments Co. Inc.
800
·
367
·
8424 sales
713
·
688
·
9345 tech
713
·
688
·
8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
·
41
·
925
·
6200 phone
Int + 41
·
41
·
925
·
6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact:
Vista de pagina 0
1 2 3 4 5 6 ... 24 25

Indice de contenidos

Pagina 1 - Instruction Manual

Valco Instruments Co. Inc.tcd2.p65Rev 4/12Printed in USAMicrovolume ThermalConductivity DetectorInstruction ManualValco Instruments Co. Inc.800 · 367·

Pagina 2

6MAINS POWER connectorFor connection to 115/230 VAC source.CHART OUTPUT connectorNormally connected to a strip chart recorder. This output has an att

Pagina 3 - Table of Contents

7System RequirementsComponents Not Included with the Detector System •Carrier gas (99.999% purity is recommended) •Ultra high purity grade gas press

Pagina 4

82 13⁄16"5 5⁄8"1 1⁄4"3"6"1 1⁄2"1"Figure 5: Detector assembly mounting dimensionsHowever, while nitrogen as a carri

Pagina 5 - Introduction

9CHANNELAVENTCHANNELBVENTCHANNELAINLET(gold ferrule)CHANNELBINLET(stainless ferrule)InstallationThe detector is usually mounted on top of the GC colum

Pagina 6 - Safety Notes and Information

10Gas ConnectionsRemember these three points discussed earlier: (1) all surfaces that contactthe gas stream must be glass or stainless steel; (2) do n

Pagina 7 - Specifications

11Installing and Purging the Gas Regulator1. Make sure the on/off valve on the helium cylinder is completely closed.Screw the CGA fitting nut of the r

Pagina 8

12InstallationColumn ConnectionTo prevent detector contamination, we strongly recom-mend disconnecting the column from the detector duringcolumn bakeo

Pagina 9

13InstallationEZU1Used with 1/16" columnEZRU21Used with 1/8" columnZU1C with fused silica adapterColumn size.32 mm ID x .5 mm OD.25 mm ID x

Pagina 10 - SERIAL NO.MODEL NO

14InstallationCHARTAB A-BOUTPUTSDETECTORSERIAL NO.Valco Instruments Co. Inc.MODEL NO.Valco Instruments Co. Inc.220VATTENUATED STRIPCHART OUTPUTCABL

Pagina 11 - System Requirements

15InstallationInitial Power-UpSince the TCD is a concentration-dependent detector, the lower the flowrate through the detector, the higher the sensiti

Pagina 12 - GC Column Selection

This page intentionally left blank for printing purposes

Pagina 13 - Installation

16InstallationThe test chromatogram accompanying your TCD was obtained under thefollowing conditions:Sample: ~100 ppm blend in Helium balanceSample vo

Pagina 14 - Gas Connections

17TroubleshootingTroubleshooting ChartProblem Possible Cause Refer to page:No signal Main or filament power switch off 4Filament temperature too low 4

Pagina 15

18Detector Fault DeterminationElectrical faults in the detector are most easily discovered by makingresistance measurements between pins of the detect

Pagina 16 - Column Connection

19MaintenanceCleaning the EnclosureThe exterior of the instrument should be cleaned regularly with a dustingbrush. If necessary, the casing can be cle

Pagina 17 - Electrical Connections

20WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Pagina 18 - MODEL NO

21Detector Performance LogDateDetector Model: Serial Number: Date of purchase:Operator Filamenttemp(°C)Detectortemp(°C)DetectorflowNoiselevelSampl

Pagina 19 - Initial Power-Up

Table of ContentsIntroductionDescription and Operating Principle ... 1Safety Notes and Inf

Pagina 20 - Single filament mode

This page intentionally left blank for printing purposes

Pagina 21 - Troubleshooting

1IntroductionDescription and Operating PrincipleThe Thermal Conductivity Detector (TCD) has been one of the most popularGC detectors since the 1950’s,

Pagina 22 - Detector Fault Determination

2Safety Notes and InformationSymbolsHOT SURFACEThe surface of the detector body may be hot while inoperation (possibly in excess of 250°C). Observe c

Pagina 23 - Maintenance

3Components of the Detector SystemsComponents of the detector system are listed in Tables 1 and 2. Check thecontents of the packages to verify that e

Pagina 24 - Warranty

4Description of Controls and ConnectorsControls and connectors are indicated in Figures 2 and 4.MAINS switch (rear panel)Controls mains (line) voltag

Pagina 25 - Detector Performance Log

5IntroductionFigure 2: Front panel controlsFigure 3: Actual filament temperature vs. filament temperature knob settingsACTUAL TEMPERATURE IN °CFILAME